Spectro UV-2510TS spectrophotometer offers high performance and reliability, which can be used in various applications.
Spectro UV-2510TS can be used extensively for qualitative and quantitative analysis in such fields as clinical analysis, medical laboratories, petro-chemistry laboratories, chemistry and biochemistry laboratories, educational labs, research laboratories, analytical laboratories, industrial testing, environmental control, water management, food processing, agriculture and water testing.
Spectro UV-2510TS has excellent baseline stability and high resolution. This spectrophotometer has 4 cell holders with a length from 5-100mm to test 4 samples.
Spectro UV-Vis 2510TS permits the instrument to be linked to a computer and a printer to display the photometric and spectral data on the PC monitor. The RS-232C interface, the port, and the included UV-VIS software link the spectrophotometer and the PC, which are compatible with Windows Platforms. The advanced 2-way communication system allows the user to provide instructions right from the computer and gives the user the ability to print and record results in an easy to use interface. In addition to saving data, the Spectro’s software can save parameters, set wavelengths and allow automatic processing of concentration.
This Spectro has 2 tungsten halogen lamps with a 2000 hour lifetime and a D2 Deuterium lamp. It has 4 multiple cell holders.
This Spectro can be used by itself or linked to a PC.
Features
- Scanning system
- User-friendly interface
- Touch screen
- Automatic lamp changing
- Automatic calculations
- Automatic wavelength settings
- 5.6 inch Color touch-screen LCD
- 5-100mm cell holder
- USB port
- Internal printer optional
- Software optional
- Lamp lifetime protection
- Power shut-down test data protection
Specifications
Optical System: |
Single Beam Spectrophotometer Scanning System |
Monochromator: |
1200 lines/mm holographic grating
C-T monochromator |
Wavelength: |
190-1100 nm |
Wavelength Accuracy: |
± 1 nm |
Wavelength Repeatability: |
≤0.5nm |
Spectral Bandwidth: |
2 nm |
Transmittance Accuracy: |
± 0.5% (T) (NBS930D) |
Stray Light: |
0.1% (T) (220nm, NaI; 360nm, NaNO2) |
Transmittance Repeatability: |
≤0.2% (T) |
Transmittance Measuring Range: |
0.0%-200.0% (T) |
Absorbance Measuring Range: |
-0.301~4.000 (A) |
Concentration Range: |
0.000~9999 (C) |
Baseline: |
± 0.003Abs (210-990nm, S model only) |
Scanning Speed: |
About 500nm/min, with a high speed 1nm step |
Noise: |
100%T≤0.2%(T) 0%T≤0.1%(T) |
Drifting: |
≤0.002%(A)/0.5h (at 250nm and 500nm, warm up 2h) |
Weight: |
20 kg |
|
Included Accessories
- 4 square optical cells 10mm.
- 2 square quartz cells 10 mm with lid
- Instruction manual
- Spectro Software
- Software Operation Manual
- Power cable
- PC cable
|